Total hip arthroplasty
(THA)is a cost effective surgical procedure for relief of pain and restoration
of function of the diseased hip. Successful THA has contributed to enhanced
mobility and comfortable independent living for people who would other
wise be substantially disabled. Currently, the majority of devices include
a bearing that consists of a metallic modular head articulating against
ultra-high molecular weight polyethylene (UHMWPE). A consequence of the
metal-on-polyethylene articulation is surface wear, liberating polyethylene
and metal wear debris. Today, osteolysis as a result wear debris is the
major cause of long-term failure of primary THA.
The increasing volume
of THA performed in younger patients, in conjunction with a greater patient
longevity, has raised expectations of implant survivorship beyond that
expected of traditional bearings. Twenty-five years ago, DeLee and Charnley
published their initial results attempting to quantify the rate of penetration
of the femoral head into a cemented, all-polyethylene acetabular component.
1 Since
this report, much has been written on the result of wear and osteolysis,
in which metal-on-polyethylene articulation is considered "the weak
link "in primary THA. 2-8 Therefore, enhancing the metal-on-polyethylene
articulation has been aggressively pursued by surgeons and industry.
Alternatively, ceramic-on-ceramic
bearings offer the opportunity to eliminate the metal-on-polyethylene
bearing and polyethylene wear debris entirely. Lyman-Smith is often credited
with the earliest published report (1963) of the use and the body's tolerance
of implanted ceramic materials. This was a porous ceramic (alumina ÐAl
2 O 3 ) imbedded with epoxy resins. 9,
10 However, ceramic material for implantation were reported thirty
years earlier in the German patent literature. 11
The advantages of ceramic components for the articulation in primary
THA include their relatively inert properties in body fluids, negligible
amounts of toxic degradation products, and their resistance to wear. Concerns
regarding ceramics frequently involve their brittle nature and mechanical
properties in high load applications. The purpose of this article is to
summarize current issues relating to the manufacturing and use of ceramic-on-ceramic
articulations for THA, including the author's clinical experience.
Unlike metals (solid solutions of elements), or polymers (long chain molecules
of carbon and hydrogen), ceramics are solid compounds of metal and non-metals.
The formation of a ceramic compound between a base metal (aluminum)and
oxygen produces a stable oxide compound in its highest state of oxidation,
in which further spontaneous reaction is not possible. Therefore, ceramics,
such as aluminum oxide (alumina - Al 2 O 3 )are inert of biological, chemical
and electrochemical reaction within the body. The subsequent strong atomic
bonds result in an inert, stiff and hard compound. 12
All articulations
in THA generate wear debris , including the ceramic-on-ceramic articulation.
However, the amount of cytokines in aseptically loose ceramic-on-ceramic
THA components is significantly less than that found in the presence of
UHMWPE debris. 13
Therefore, based on the particle size, volume and bioinert properties,
the incidence of wear debris induced osteolysis in THA with ceramic-on-ceramic
articulations may be less than that seen with standard metalon-polyethylene
articulations.
Alumina oxide ceramics is manufactured from purified natural minerals.
Pure oxide is milled to a mesh of 0. 1 10 µm. Dies that are manufactured
to the component shape are used in which the oxide power is compressed
using 70 Mpa pressure (~10, 000 psi). The compressed product is transferred
to an oven and baked at low temperature to its point of fusion (~1, 700
degrees Fahrenheit), and is slowly cooled to room temperature. On the
MOHS hardness scale, current alumina products for THA are classed 9. The
hardest material rated just above alumina, class 10, is diamond. 9
Current ceramic components have a reported overall incidence of fracture
five times less than early generation ceramic components (early Ð 1:2,
000 current Ð1:10, 000). This is a result of manufacturing processes that
include a smaller grain size (1. 8 µversus 4. 5 µ), increased density
(3. 98 g/cc versus 3. 50 g/cc), and a higher alumina purity (99. 9%versus
99. 5%). These variables have led to optimized component material properties.
|
|
|
|
Figure
1. Ceramic-ceramic total hip arthroplasty |
|
MATERIALS
AND METHODS
Between June 1997 and February 2001, 75 THA's (75 hips in 69 patients).
The acetabular component used in all cases was a press-fit, porous-coated
titanium shell (Transcend TM Cup ÐWright Medical Technology, Memphis,
TN). An alumina ceramic acetabular bearing was inserted into the metal
shell and locked with an 18 deg. taper. The alumina acetabular bearing
articulates with a modular alumina femoral head component
(Figure 1). All acetabulae were under-reamed by 1 mm to ensure
stability upon impaction. Screws to augment acetabular fixation were generally
not used. All patients were followed prospectively as part of an FDA/IDE
approved study. The patients were evaluated clinically and radiographically
pre-operatively, and post-operatively at 3, 6 and 12 months and annually
thereafter. Outcome evaluation included Harris hips scores and SF-12 questionnaires.
There were 28 (37%)left
hips, 33 (44%)right hips and 14 (19%)bilateral hips. 40 (53%)were male
and 35 (47%)were female. Mean patient age at surgery was 48 years (SD
=13, range 18-75 years). Pre-operative diagnoses included osteoarthritis
45 (60%), developmental dysplatic hip 21 (28%), avascular necrosis 6 (8%),
traumatic arthritis 2 (3%), Legg-Calve-Perthes disease 1 (1%). Mean ±SD
follow-up time for all cases was 11 ±12 months.
RESULTS
To date, all components are well fixed. There have been no cases of ceramic
head or liner fracture postoperatively and no signs of grossly visible
wear. The acetabular components were fixed without screws in all but two
hips (one patient). The acetabular component abduction angle averaged
43 degrees (SD =4 degrees, range 30-53 degrees). Intraoperatively, one
femur cracked during component insertion and was treated by cerclage wires.
One ceramic liner chipped when inserted eccentrically and was replaced
at the time. Postoperatively, one hip underwent irrigation and debridement
3 weeks postoperatively for increased pain and low-grade temperature and
was culture negative. One liner was mal-seated and replaced on post-op
day 3. There were no dislocations or infections.
|
|
|
|
|
|
|
Figure
2. Ceramic-ceramic bearings may be especially advantageous
in dysplastic hips with small sockets where polyethylene wear due
to thin plastic liners remains a serious problem. |
DISCUSSION
Early results have demonstrated no major problems associated with ceramic-ceramic
bearings in THA, such as bearing fracture or catastrophic wear. All preliminary
series of ceramicceramic bearings have shown an incidence of liner chipping
or mal-seating, especially early on. In fact, ceramic liners are generally
easier to insert than polyethylene liners but need to be properly seated
by hand before impaction. Since, unlike polyethylene, bearing thickness
does not affect wear, ceramic-ceramic bearings are especially useful for
dysplastic hips (Figure 2). These patients
are typically young, with small socket components, where thin polyethylene
typically leads to a high rate of polyethylene wear.
In an FDA approved,
multi-center, clinical trial of 333 THA's with the same component (of
which the author has participated and results are combined), Garino reported
shortterm results. 14
As with the current study, the preliminary results had no post-operative
cases of ceramic bearing fracture or visible wear. There were 2 additional
intraoperative liner "chip " fractures due to eccentric insertion
treated by intraoperative exchange of the liner.
Similarly, Bizot,
et al, reported a 93%survivorship rate in 234 THA's with a non-cemented,
metal-backed, ceramic-on ceramic acetabular component, when revision was
the endpoint. 15
Of these cases, only one was ceramic related (fractured femoral head).
There were three hips in which the acetabular component had a complete,
non-progressive radiolucent line less than 1 mm thick. In each case, neither
component migration nor osteolysis adjacent to the acetabular component
was detected radiographically. These intermediate reports of this component
show results equal to that of conventional THA with metal-on-polyethylene
articulation.
CONCLUSION
Early results of the use of ceramic-ceramic bearings for THA show promise
for eliminating polyethylene debris and reducing wear debris. Since polyethylene
appears to induce a stronger histological reaction than inert ceramic
wear particles, there is potential for less osteolysis even given an equivalent
concentration of wear particles. Ceramic bearings may prove especially
useful in younger patients and small, dysplastic hips.
Stephen B. Murphy,
MD is a Clinical Instructor in Orthopaedic Surgery at Harvard Medical
School.
Wael K. Barsoum,
MD is a Total Joint Arthroplasty Fellow at the New England Baptist
Hospital, Boston, MA.
Address correspondence
to:
Stephen B. Murphy,
MD
New England
Baptist Hospital
Department of Orthopaedic Surgery
125 Parker Hill Avenue
Boston, MA 02120
|